MicroBio Sampling Heads

All sampler kits are supplied with one head type specified when ordering. However, additional heads can be ordered at the same time or later when required.

Model220x1.0 s/s400x0.7 a/a400x0.7 s/s400x1.0 s/s (HiFlow)
Holes220400400400
Hole diameter1.0 mm
0.7 mm0.7 mm1.0 mm
Flow rate100 L/min100 L/min100 L/min180 L/min
Plate size compatibility90 mm Petri dish
55mm contact plate
90 mm Petri dish90 mm Petri dish90 mm Petri dish
d50 particle cut-off1.7 µm1.35 µm1.35 µm1.71 µm
Particle velocity9.65 m/s10.83 m/s10.83 m/s9.55 m/s
Material316L stainless steelAluminium316L stainless steel316L stainless steel
FinishElectropolishAnodisedElectropolishElectropolish
ISO14698 Tested Design
Autoclave safe
Part NumberA-00021A-00020A-00670A-00183

MicroBio air samplers are calibrated to use a particular type of head, so when ordering more please ensure the same type is ordered.

MicroBio media support plate dust cover.

This accessory protects the media support plate and blower inlet of MicroBio bioaerosol samplers from contamination during transport and setup.  It’s especially useful for safeguarding MicroBio MB2-RSH remote and fixed head units in isolator cabinets when not in use.

This autoclave-safe 316 grade stainless steel cover is suitable for all MicroBio Bioaerosol sampler models. 

You can purchase this individually with part number A-00771 or order it as part of a MicroBio air sampler kit.  Simply add ‘-DC’ to the kit part number, for example A-00024-S-220-55-DC.